B. BRAUN MEDICAL INC. INFUSOMAT; SET, ADMINISTRATION, INTRA Back to Search Results
Model Number 363420
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/15/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).It was confirmed that the device involved will not be available for evaluation, the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: event 1: it was reported that during chemotherapy of taxol, the filter leaked.No injury reported.
 
Additional Manufacturer Narrative
Event 1: this report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample was returned for evaluation.Visual examination of the sample was unable to note any defects.The sample was leak tested per specification and liquid was noted to be coming from the bottom vent hole in the air eliminating filter.Based on the evaluation of the sample the reported defect is confirmed.Based on the data from the investigation we are unable to determine the root cause of the reported incident.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOMAT
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
MDR Report Key9695518
Report Number2523676-2020-00017
Device Sequence Number1809938
Product Code FPA
UDI-Device Identifier04046964294594
UDI-Public04046964294594
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K142036
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number363420
Device Catalogue Number363420
Device Lot Number0061691726
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2020
Initial Date Received by Manufacturer 01/15/2020
Supplement Date Received by Manufacturer01/15/2020
Initial Report FDA Received Date02/11/2020
Supplement Report FDA Received Date05/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-