| Model Number |
402808 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Complete Heart Block (2627)
|
| Date of Event |
01/13/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An event of heart block was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported heart block could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference: 2184149-2020-00017.During a slow pathway modification for an av-nodal reentrant tachycardia (avnrt), heart block occurred.After an rf session was being delivered, it was noticed the catheter jumped out of position and the generator was requested to be called off.The rf was not ceased until 4-5 seconds after the request.Av block occurred due to the delayed response in turning off the generator.The patient required a permanent pacemaker implant and was in a stable condition.
|
| |
|
Event or Problem Description
|
|
The catheter moving out of positive was caused by the handling of the catheter.
|
| |
|