| Model Number |
502-03-50D |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Injury (2348); Inadequate Osseointegration (2646)
|
| Date of Event |
01/14/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.Not returned.
|
| |
|
Event or Problem Description
|
|
Tritanium primary cluster cup revised for loosening on (b)(6) 2020.Though four screws were listed in the original usage report, it was not noted which were wasted.Two screws were in the patient at time of revision--both removed, one broken.Restoration modular body was also exchanged at time of revision.Patient's right hip was revised to competitor (acetabular) and stryker (femoral) devices.Update per response: "no allegations were made against any implants other than the cup.".
|
| |
|
Search Alerts/Recalls
|