Model Number 71674034
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/24/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure, the surgeon was performing an intertan nail and when he used the 7.0 compression screw drill, the tip broke off.The pieces were retrieved and not left in the patient.
 
Additional Manufacturer Narrative
Additional info: d4, b2, g7, g9, h2, h3, h4, h6, h10.Correction: d4.Results of investigation: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The pilot tip of the drilling end on the device fractured off and was not returned.The device was manufactured in 2015 and shows signs of extensive wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7.0MM COMPRESSION SCREW DRILL
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9697060
Report Number1020279-2020-00513
Device Sequence Number11487474
Product Code KTT
UDI-Device Identifier03596010560544
UDI-Public03596010560544
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/04/2025
Device Model Number71674034
Device Catalogue Number71674570
Device Lot Number15FNG0023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2020
Initial Date Received by Manufacturer 01/27/2020
Supplement Date Received by Manufacturer01/27/2020
Initial Report FDA Received Date02/11/2020
Supplement Report FDA Received Date04/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-