Model Number MMT-523LNAL
Medical Device Problem Code Failure to Cycle (1142)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had a compromised force sensor system alarm.The customer stated that they were able to clear the alarm and rewind the insulin pump.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the displacement test but a33 alarm during the basic occlusion test due to loose drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery test due to a33 alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9698722
Report Number3004209178-2020-68574
Device Sequence Number14142221
Product Code OYC
UDI-Device Identifier00643169151291
UDI-Public(01)00643169151291
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAL
Device Catalogue NumberMMT-523LNAL
Device Lot NumberA2523LNALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2020
Initial Date Received by Manufacturer 02/09/2020
Supplement Date Received by Manufacturer04/17/2020
Initial Report FDA Received Date02/12/2020
Supplement Report FDA Received Date04/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age24 YR
Patient Weight64
-
-