• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number TRM-210
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/10/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Physician reported right side rupture.The device has been explanted.
 
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Device evaluation: visual analysis of the returned device identified: underweight, broken shell, wear abrasion, crease fold, missing, non-penetrating nicks.A microscopic analysis was performed which identify sharp edge and with non-penetrating nicks assessed as surgical impact opening on posterior side.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a sharp edge opening with non-penetrating nicks assessed as surgical impact.Non-penetrating nicks.
 
Event or Problem Description
Physician reported right side rupture.The device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key9700248
Report Number9617229-2020-01145
Device Sequence Number19268150
Product Code FTR
UDI-Device Identifier10888628008038
UDI-Public10888628008038
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2020
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/08/2013
Device Model NumberTRM-210
Device Catalogue NumberTRM210
Device Lot Number1662398
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/29/2020
Initial Date Received by Manufacturer 01/14/2020
Supplement Date Received by Manufacturer02/29/2020
Initial Report FDA Received Date02/12/2020
Supplement Report FDA Received Date03/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
-
-