| Model Number |
TRM-210 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/10/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Physician reported right side rupture.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Device evaluation: visual analysis of the returned device identified: underweight, broken shell, wear abrasion, crease fold, missing, non-penetrating nicks.A microscopic analysis was performed which identify sharp edge and with non-penetrating nicks assessed as surgical impact opening on posterior side.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a sharp edge opening with non-penetrating nicks assessed as surgical impact.Non-penetrating nicks.
|
| |
|
Event or Problem Description
|
|
Physician reported right side rupture.The device has been explanted.
|
| |
|