Model Number 357469
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/20/2020
Type of Reportable Event Malfunction
Event or Problem Description
A passeo-18 peripheral dilation catheter was selected for treatment.During the procedure the balloon burst.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon has burst longitudinal over a length of about 17 mm.Microscopic inspection of the balloon showed several deep scratches in close vicinity of the tear.It seems likely that the tear in the balloon was caused by a hard, sharpedged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no material or manufacturing related root cause was determined.The root cause is most likely related to the patients anatomy (i.E.Calcified lesion).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-18 2.5/80/90
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key9700315
Report Number1028232-2020-00714
Device Sequence Number2860299
Product Code LIT
UDI-Device Identifier07640119558806
UDI-Public07640119558806
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072765
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/30/2022
Device Model Number357469
Device Catalogue NumberSEE MODEL NO.
Device Lot Number11192236
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2020
Initial Date Received by Manufacturer 02/11/2020
Supplement Date Received by Manufacturer04/20/2020
Initial Report FDA Received Date02/12/2020
Supplement Report FDA Received Date04/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-