• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD CARELINK ENCORE PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD CARELINK ENCORE PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 29901A
Device Problems Failure to Power Up (1476); Battery Problem (2885); Power Problem (3010); Unexpected Shutdown (4019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/15/2020
Event Type  malfunction  
Manufacturer Narrative
At analysis it was noted that the programmer's battery was dead and would no longer charge.The device powered on with a good power supply and charged a new battery.No power supply was returned with the programmer.Its battery was replaced, the software was reloaded and updated and it then passed all safety, console and systems tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer 'flickered' on when powered on but powered off shortly after.Trouble-shooting discussion included removing the battery to plug in to power and if issue persists to return the programmer for service.It was additionally suggested that the user could try ordering a new battery and trying that.The status of the programmer is currently unknown.There were no patient complications associated with the event.It was further reported that the programmer was not powering on and was returned for service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK ENCORE PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9702097
MDR Text Key179249182
Report Number3004593495-2020-00141
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00643169763548
UDI-Public00643169763548
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number29901A
Device Catalogue Number29901A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2020
Date Manufacturer Received01/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-