| Model Number |
X SERIES |
| Medical Device Problem Code |
Reset Problem (3019)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
01/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to monitor a patient (age & gender unknown), the device restart/reboot itself multiple times.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported malfunction was observed during review of the device activity logs.However, the device was put through extensive internal and external testing without duplicating the report.The device was recertified and returned to the customer.The multifunction cable for the device was proactively replaced.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|