MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #6 13MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON CS INSERT #6 13MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5531-G-613
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Injury (2348)
Date of Event 01/15/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.The following devices were also listed in this report: triathlon p/a cr beaded #6l; cat#: 5517f601; lot#: aan3s1; tritanium bplate triathlon s6; cat# 5536b600; lot# ctd7479; tritanium patella-asymmetric; cat# 5552-l-350; lot# a6jt.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Not returned.
 
Event or Problem Description
As reported: "left knee revision.Revision done due to infection.All implants were explanted.".
 
Additional Manufacturer Narrative
Reported event: an event regarding infection involving triathlon insert was reported.The event was not confirmed.Method & results: -device evaluation and results: not performed as product was not returned.-clinician review: no medical records were received for review with a clinical consultant.-device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot or sterile lot.Conclusions: infection is a known possible adverse outcome of surgery, as noted in the instructions for use.All stryker products sold as sterile are validated to a minimum sterility assurance level (sal) of 10^-6 in accordance to applicable iso standards.When the sterile barrier of any device is opened, the sterility of that device becomes a function of handling and surgical technique and is beyond stryker¿s control.Infection due to bacterial contamination, as supplied from the manufacturer, is an extremely rare event (see technical report 0000128012).The exact cause of the event could not be determined because insufficient information was provided.Additional information including pathology reports, operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.Product surveillance will continue to monitor for trends.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.
 
Event or Problem Description
As reported: "left knee revision.Revision done due to infection.All implants were explanted.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INSERT #6 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9702562
Report Number0002249697-2020-00278
Device Sequence Number14024365
Product Code MBH
UDI-Device Identifier07613327045949
UDI-Public07613327045949
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2017
Device Model Number5531-G-613
Device Catalogue Number5531G613
Device Lot NumberLDD822
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/15/2020
Supplement Date Received by Manufacturer02/27/2020
Initial Report FDA Received Date02/12/2020
Supplement Report FDA Received Date03/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
Patient Weight118
-
-