| Model Number |
MMT-722NAB |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Insulin pump passed the rewind, basic occlusion, prime and displacement test.Pump received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had and intermittent failure that was not detected during our testing.Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated reported that the insulin pump had motor error alarm once and a no delivery alarm during bolus.Customer reports the drive support cap appears normal.Customer was able to complete insulin pump rewind.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|