The incident was noticed during inspection of the rotaflow drive.The affected rotaflow drive was shipped to germany for repair by the manufacturer emtec under rma#40385.Incoming goods in rastatt (b)(6)2020 functionaltest in rastatt (b)(6)2020 failure not confirmed sent to emtec (b)(6)2020 back from emtec(b)(6)2020 outgoing goods (b)(6)2020 according to the service report rma2020-10052 from emtec dated on (b)(6)2020 functional test performed to reproduce the reported failure.The drive was connected to the test console by emtec and test started.The console was started in the stand alone mode.The speed increases so far to be able to trigger a bubble alarm in the test circuit.The tube is narrowed a little so that bubbles are activated, the drive stops with a bubble alarm.Carry out the experiment several times, always with a bubble alarm.Failure could not be reproduced/confirmed.About the described missing pin.By version of the serial number (b)(6) it is normal that pin 8 is missing and this has no influences of the functionalty the rotaflow drive.Newer versions of drives have a blind pin.Drive passed all tests.The rotaflow drive was tested in rastatt according to the service protocol (see service report 10478149 dated on (b)(6)2020).The rotaflow risk analysis version v06 (dms# 2023689) chapter h1.1.1.18 was reviewed on (b)(6)2020 with the following outcome: the most possible causes for the reported failure "bubble sensor doesn´t work" could be determined as: bubble/flow sensor failure, e.G.: * malfunction of bubble/flow sensor electronics * dried contact gel * user forgot renewing contact gel * mechanical defects (impact) * sensor not detected although sensor is connected * device used out of specification * software error the reported failure "bubble sensor doesn´t work" occurred during inspection and could not be confirmed.The device was directly involved in the incident.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
|