• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AD-TECH MEDICAL INSTRUMENT CORP. CRANIAL DRILL BITS AND ACCESSORIES; DRILL SLEEVE GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AD-TECH MEDICAL INSTRUMENT CORP. CRANIAL DRILL BITS AND ACCESSORIES; DRILL SLEEVE GUIDE Back to Search Results
Model Number DSG-6.3-090-2.4N
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mdr 2183456-2019-00026 has been filed past the 30-day timeframe.Through the 2019 fda inspection, it was identified that while ad-tech was evaluating reportability for actual events in which death or serious injury had occurred, ad-tech had failed to submit events in which a malfunction was reported and could result in a serious injury if the event were to recur.As part of the investigation, a two (2) year retrospective review of complaints was performed to determine which complaints required submission of an mdr.Mdr-2183456-2019-00026 was submitted as a conservative measure due to the recalls tied to this issue.
 
Event or Problem Description
An ad-tech clinical specialist attending a case at a user facility reported that the drill bit seized in the drill sleeve guide, possibly due to overtightening the drill stop, bending the bit.The second bit was used and the case proceeded without patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRANIAL DRILL BITS AND ACCESSORIES
Common Device Name
DRILL SLEEVE GUIDE
Manufacturer (Section D)
AD-TECH MEDICAL INSTRUMENT CORP.
400 west oakview parkway
oak creek, wi
MDR Report Key9705921
Report Number2183456-2019-00026
Device Sequence Number3474561
Product Code HBE
UDI-Device Identifier00841823106520
UDI-Public(01)00841823106520(17)191201(10)0116389712121
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K170442
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 02/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2019
Device Model NumberDSG-6.3-090-2.4N
Device Catalogue NumberDSG-6.3-090-2.4N
Device Lot Number712121
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/02/2018
Initial Report FDA Received Date02/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2183456-08212019-R-004
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-