| Model Number |
DSG-6.3-090-2.4N |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Mdr 2183456-2019-00026 has been filed past the 30-day timeframe.Through the 2019 fda inspection, it was identified that while ad-tech was evaluating reportability for actual events in which death or serious injury had occurred, ad-tech had failed to submit events in which a malfunction was reported and could result in a serious injury if the event were to recur.As part of the investigation, a two (2) year retrospective review of complaints was performed to determine which complaints required submission of an mdr.Mdr-2183456-2019-00026 was submitted as a conservative measure due to the recalls tied to this issue.
|
| |
|
Event or Problem Description
|
|
An ad-tech clinical specialist attending a case at a user facility reported that the drill bit seized in the drill sleeve guide, possibly due to overtightening the drill stop, bending the bit.The second bit was used and the case proceeded without patient harm.
|
| |
|
Search Alerts/Recalls
|