| Model Number |
TRF-605 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/16/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iii.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.Device evaluation: visual analysis of the returned device identified crease fold, deformation, and cloudiness/bubbles in the gel observed after autoclave disinfection process.Weight measurement of the device identified device weight was within specification.Based on the device analysis the final assessment was no issues found related with the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Healthcare professional additionally reported "patient's concern with the product".Device has been explanted.
|
| |
|