Model Number TSX-470
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 01/15/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
 
Event or Problem Description
Healthcare professional reported right side capsular contracture, baker grade iii.Device remains implanted.
 
Additional Manufacturer Narrative
Device evaluation: the device related to the reported event of capsular contracture was received on february 29, 2020 with lot number 3019449.Analysis of the returned device identified: weight to the spec, crease fold, deformation, yellow biological tissue.Cloudy and voids in the gel observed after autoclave cycle.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.
 
Event or Problem Description
Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key9706175
Report Number9617229-2020-01477
Device Sequence Number2396437
Product Code FTR
UDI-Device Identifier10888628034365
UDI-Public10888628034365
Combination Product (Y/N)N
Initial Reporter StateCA
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/06/2022
Device Model NumberTSX-470
Device Catalogue NumberTSX-470
Device Lot Number3019449
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/29/2020
Initial Date Received by Manufacturer 01/16/2020
Supplement Date Received by Manufacturer02/29/2020
Initial Report FDA Received Date02/13/2020
Supplement Report FDA Received Date03/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age41 YR
-
-