• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL Back to Search Results
Catalog Number PWF030
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Irritation (1941); Pain (1994); Swelling (2091); Urinary Tract Infection (2120)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the customer experienced a uti with pain, irritation and swelling with use of the pure wick.Patient was hospitalized and given antibiotics for the uti.
 
Additional Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.A potential root cause for this failure could be "inadequate material characteristics, degradation over time".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "caution: this product contains dry natural rubber contraindications: ¿ patients with urinary retention warnings: ¿ do not use the purewicktm female external catheter with bedpan or any material that does not allow for sufficient airflow.¿ to avoid potential skin injury, never push or pull the purewicktm female external catheter against the skin during placement or removal.¿ never insert the purewicktm female external catheter into vagina, anal canal, or other body cavities.¿ discontinue use if an allergic reaction occurs.¿ after use, this product may be a potential biohazard.Dispose of in accordance with applicable local, state and federal laws and regulations.Precautions: ¿ not recommended for patients who are: ¿ agitated, combative, or uncooperative and might remove the purewicktm female external catheter ¿ having frequent episodes of bowel incontinence without a fecal management system in place ¿ experiencing skin irritation or breakdown at the site ¿ experiencing moderate/heavy menstruation and cannot use a tampon ¿ do not use barrier cream on the perineum when using the purewicktm female external catheter.Barrier cream may impede suction.¿ not recommended for use on patients with a known latex allergy.¿ proceed with caution in patients who have undergone recent surgery of the external urogenital tract.¿ always assess skin for compromise and perform perineal care prior to placement of a new purewicktm female external catheter.¿ maintain suction until the purewicktm female external catheter is fully removed from the patient to avoid urine backflow.Recommendations: ¿ perform each step with clean technique.In the home setting, wash hands thoroughly before device placement.¿ prior to connecting the purewicktm female external catheter to hospital wall suction tubing, verify suction function by covering the open end of the suction tubing with one hand and observing the pressure dial.If the pressure does not increase when the line is covered, verify that the tubing is secured, connected, and not kinked.¿ ensure the purewicktm female external catheter remains in the correct position after turning the patient.Remove the purewicktm female external catheter prior to ambulation.¿ properly placing the purewicktm female external catheter snugly between the labia and gluteus holds the purewicktm female external catheter in place for most patients.Mesh underwear may be useful for securing the purewicktm female external catheter for some patients.¿ assess device placement and patient¿s skin at least every 2 hours.¿ replace the purewicktm female external catheter every 8-12 hours or when soiled".H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the customer experienced a uti with pain, irritation and swelling with use of the pure wick.The patient was hospitalized and given antibiotics for the uti.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREWICK FEMALE EXTERNAL
Common Device Name
PUREWICK
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9706681
Report Number1018233-2020-01003
Device Sequence Number3341779
Product Code NZU
UDI-Device Identifier00801741143083
UDI-Public(01)00801741143083
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberPWF030
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/21/2020
Supplement Date Received by Manufacturer03/15/2020
Initial Report FDA Received Date02/13/2020
Supplement Report FDA Received Date03/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-