| Model Number |
ADVIA 1800 CHEMISTRY - LIPASE (LIP) |
| Medical Device Problem Code |
Imprecision (1307)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Siemens healthcare diagnostics, inc.Has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using advia chemistry lipase reagent lot 485700.Preliminary investigation has indicated that not all cartons within this reagent lot are impacted.An urgent medical device recall (umdr) chc 20-03.A.Us was sent to us customers and an urgent field safety notice (ufsn) chc 20-03.A.Ous was sent to ous customers in december 2019.The umdr and ufsn advise customers of the investigation with the advia chemistry lipase reagent lot 485700 and provide instructions to help determine if the lipase reagent cartons in their inventory are impacted.If the cartons in the customers inventory are not impacted, the customers are instructed to proceed to using the cartons according to the instructions for use.If the cartons in the customers inventory are impacted, the customers are instructed to discard the impacted cartons and can request replacement.The next reagent lot is expected to be available by february 2020.If replacement kits are not available, customers are instructed to test patient samples using an alternate methodology.The customer received umdr chc 20-03.A.Us and followed the instructions in the bulletin.The customer resumed lip testing with acceptable quality control results.The instrument is performing within specification.No further evaluation of the device is required.
|
| |
|
Event or Problem Description
|
|
The customer obtained imprecise lipase (lip) patient and quality control (qc) results while using advia 1800 chemistry.The initial patient results were not reported to the physician(s).The samples were repeated using an alternate advia 1800 chemistry.The repeat results were reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely depressed lip results.
|
| |
|
Search Alerts/Recalls
|