Model Number DDK2-2.4-30X
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/31/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mdr 2183456-2019-00020 has been filed past the 30-day timeframe.Through the 2019 fda inspection, it was identified that while ad-tech was evaluating reportability for actual events in which death or serious injury had occurred, ad-tech had failed to submit events in which a malfunction was reported and could result in a serious injury if the event were to recur.As part of the investigation, a two (2) year retrospective review of complaints was performed to determine which complaints required submission of an mdr.Mdr-2183456-2019-00020 was submitted as a conservative measure due to the recalls tied to this issue.
 
Event or Problem Description
On (b)(6) 2019 ad-tech received a complaint from a user facility regarding the fusing of an ad-tech drill bit to a rosa guide during a seeg procedure.The procedure could not proceed and was subsequently rescheduled for the following day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRANIAL DRILL BITS AND ACCESSORIES
Common Device Name
CRANIAL DRILL BITS
Manufacturer (Section D)
AD-TECH MEDICAL INSTRUMENT CORP.
400 west oakview parkway
oak creek, wi
MDR Report Key9707293
Report Number2183456-2019-00020
Device Sequence Number19670714
Product Code HBE
UDI-Device Identifier00841823101167
UDI-Public(01)00841823101167(17)200401(10)0120156208140673
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K170442
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 02/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2020
Device Model NumberDDK2-2.4-30X
Device Catalogue NumberDDK2-2.4-30X
Device Lot Number208140673
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/31/2019
Initial Report FDA Received Date02/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2183456-08212019-R-004
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-