• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AD-TECH MEDICAL INSTRUMENT CORP. RD DEPTH ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AD-TECH MEDICAL INSTRUMENT CORP. RD DEPTH ELECTRODE Back to Search Results
Model Number RD10R-SP04X-000
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mdr 2183456-2020-00011 has been filed past the 30-day timeframe.Through the 2019 fda inspection, it was identified that while ad-tech was evaluating reportability for actual events in which death or serious injury had occurred, ad-tech had failed to submit events in which a malfunction was reported and could result in a serious injury if the event were to recur.As part of the investigation, a two (2) year retrospective review of complaints was performed to determine which complaints required submission of an mdr.
 
Event or Problem Description
On (b)(6) 2017 ad-tech received a complaint of a bent electrode prior to clinical use.There was no report of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RD DEPTH ELECTRODE
Common Device Name
DEPTH ELECTRODE
Manufacturer (Section D)
AD-TECH MEDICAL INSTRUMENT CORP.
400 west oakview parkway
oak creek, wi
MDR Report Key9707522
Report Number2183456-2020-00011
Device Sequence Number11598575
Product Code GZL
UDI-Device Identifier00841823107244
UDI-Public(01)00841823107244(17)190601(10)0109231208140645
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K163355
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2019
Device Model NumberRD10R-SP04X-000
Device Catalogue NumberRD10R-SP04X-000
Device Lot Number208140645
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2017
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2017
Initial Report FDA Received Date02/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-