• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 CORAIL2 NON COL HO SIZE 14; HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 CORAIL2 NON COL HO SIZE 14; HIP FEMORAL STEM Back to Search Results
Model Number L20314
Medical Device Problem Codes Corroded (1131); Osseointegration Problem (3003)
Health Effect - Clinical Codes Calcium Deposits/Calcification (1758); Foreign Body Reaction (1868); Pain (1994); Osteolysis (2377); Inadequate Osseointegration (2646); Test Result (2695); No Code Available (3191)
Date of Event 02/06/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).No code available use for device revision or replacement.
 
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pcf and medical records.After review of medical records, patient was revised to addressed periprosthetic osteolysis, acetabular component loosening and elevated metal ions.Revision notes indicated dark stained reactive synovium encountered and debrided.With torque on the prosthetic head, femoral micromotion was confirmed consistent with the diagnosed of loosening.Disassembly of the modular head revealed relatively little black corrosive debris inside the taper and the trunnion appeared normal.A long curette was used to ensure removal of any endosteal calcification potentially diverting the femoral reamers.There was no significant evidence of corrosion on the backside of the liner of the inside of the shell.The 2 acetabular fixation screws were removed.The cup loose watery region of anterior rim was overhanging so it was used to remove that rim of bone and dislodge the cup.There was a remaining reasonably hemispherical acetabular cavity.Doi: (b)(6) 2009; dor: (b)(6) 2019 right hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint(b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 NON COL HO SIZE 14
Common Device Name
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key9707624
Report Number1818910-2020-04753
Device Sequence Number16997613
Product Code KWA
UDI-Device Identifier10603295258056
UDI-Public10603295258056
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K042992
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2019
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/30/2012
Device Model NumberL20314
Device Catalogue NumberL20314
Device Lot Number2472146
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/28/2020
Supplement Date Received by Manufacturer02/14/2020
03/18/2020
Initial Report FDA Received Date02/13/2020
Supplement Report FDA Received Date03/09/2020
03/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
-
-