• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DRILL FLEXIBLE ENDOSCOPIC CANN 7MM; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DRILL FLEXIBLE ENDOSCOPIC CANN 7MM; BIT, DRILL Back to Search Results
Model Number 7209008
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/20/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during an unknown procedure, the 7mm flexible drill broke while beginning to drill the tunnel.It is unknown if a back-up device was available or if there was a delay in the scheduled surgery.No patient injuries or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10: h3,h6: one 7209008 drill flexible endoscopic cann 7mm, intended for use in treatment, has not been returned for evaluation.This is a six-year-old reusable device.Due to unavailability, investigation was limited.The information provided states: ¿during an unknown procedure, the 7mm flexible drill broke while beginning to drill the tunnel¿.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: excessive force applied, surgical techniques, proper care and handling.The instruction for use states: ¿do not use these instruments as levers for manipulating hard tissue or bone.Excessive force should not be applied to the instrument when manipulating soft tissue, bone, or hard objects.Misuse of these instruments may result in bent distal tips or jaws; and dull or uneven cutting edges¿.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.A review of complaints and manufacturing batch records was preformed, no complaints of this failure were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL FLEXIBLE ENDOSCOPIC CANN 7MM
Common Device Name
BIT, DRILL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key9708498
Report Number1219602-2020-00215
Device Sequence Number3171891
Product Code HTW
UDI-Device Identifier03596010601445
UDI-Public03596010601445
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7209008
Device Catalogue Number7209008
Device Lot Number50517092
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/20/2020
Supplement Date Received by Manufacturer03/12/2020
Initial Report FDA Received Date02/13/2020
Supplement Report FDA Received Date03/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-