• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number RONYX40026UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information: a highly calcified lesion.The devices were prepped per ifu with no issues noted.It was indicated that excessive force was used when withdrawing both onyx devices.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During a procedure an attempt was made to use one resolute onyx rx (4.00 x 26mm) and one resolute onyx rx (2.25 x 18mm) drug eluting stents to treat a lesion.The devices were inspected with no issues noted.The lesions were not pre-dilated.Negative prep was not performed.The devices did not pass through a previously deployed stent.Resistance was encountered when advancing the resolute onyx rx (4.00 x 26mm) device.Resistance was not encountered when advancing the resolute onyx rx (2.25 x 18mm) device.Excessive force was not used during delivery of either device.It was reported that both stents dislodged during removal following failed deliveries.It was reported that after the resolute onyx rx (4.00 x 26mm) failed to cross the lesion it was pulled back into the y adapter.The stent then got "hung up" on the y adapter and came off the delivery catheter.It was reported that the resolute onyx rx (2.25 x 18mm) dislodged in the y adapter.Both devices were retrieved successfully.The dislodged stents were removed by the physicians hand.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key9710349
Report Number9612164-2020-00713
Device Sequence Number3270145
Product Code NIQ
UDI-Device Identifier00643169557192
UDI-Public00643169557192
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/11/2020
Device Model NumberRONYX40026UX
Device Catalogue NumberRONYX40026UX
Device Lot Number0009233803
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/05/2020
Supplement Date Received by Manufacturer04/01/2020
Initial Report FDA Received Date02/14/2020
Supplement Report FDA Received Date04/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-