Model Number 407358
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/07/2020
Type of Reportable Event Malfunction
Event or Problem Description
The report was submitted inadvertently as this event occurred during preparation.
 
Additional Manufacturer Narrative
One 8.5f swartz braided introducer sheath and dilator were received for evaluation.Resistance was noted near the distal end of the dilator when a brk needle/stylet assembly from current inventory was advanced through the returned dilator and sheath.The dilator tubing was cut lengthwise and evidence of skiving was noted along the inner wall of the dilator.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the received condition of the device and the information provided, the cause of the skiving remains unknown.
 
Event or Problem Description
This event is being reported based on analysis revealing skiving of the dilator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, LAMP¿ 90, 63 CM LENGTH, 8.5F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
MDR Report Key9710939
Report Number3005334138-2020-00058
Device Sequence Number19746539
Product Code DYB
UDI-Device Identifier05414734205535
UDI-Public05414734205535
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K052644
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model Number407358
Device Catalogue Number407358
Device Lot Number7193353
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/14/2020
Supplement Date Received by Manufacturer02/14/2020
Initial Report FDA Received Date02/14/2020
Supplement Report FDA Received Date02/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-