| Model Number |
407358 |
| Medical Device Problem Code |
Flaked (1246)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The report was submitted inadvertently as this event occurred during preparation.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8.5f swartz braided introducer sheath and dilator were received for evaluation.Resistance was noted near the distal end of the dilator when a brk needle/stylet assembly from current inventory was advanced through the returned dilator and sheath.The dilator tubing was cut lengthwise and evidence of skiving was noted along the inner wall of the dilator.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the received condition of the device and the information provided, the cause of the skiving remains unknown.
|
| |
|
Event or Problem Description
|
|
This event is being reported based on analysis revealing skiving of the dilator.
|
| |
|