Model Number 71645030
Medical Device Problem Code Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/24/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during surgery, a screw driver did not capture the screw.It seems that the screw hole was already damaged/licking before using it.Procedure was completed using a 35mm screw.No harm to patient.No delay.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The hex head is damaged/ deformed, rendering the device inoperable.The device was manufactured in 2019.The medical investigation concluded that it was communicated that the procedure was completed with a backup device.Per complaint, no patient harm or surgical delay resulted from the event; therefore, no further medical assessment is warranted at this time.The product evaluation revealed that the screw failed the gage for hex depth.This led to the screw not being able to mate for the driver.A review of complaint history did not reveal additional complaints for the listed batch for the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do have reason to suspect that the product failed to meet product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The potential probable cause for this event is likely a manufacturing process error.Based on this investigation, the need for corrective action is indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIGEN LOW PROFILE SCREW 5.0MM X 30MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9711059
Report Number1020279-2020-00541
Device Sequence Number2500371
Product Code HSB
UDI-Device Identifier00885556215593
UDI-Public00885556215593
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111025
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,foreign,health profe
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71645030
Device Catalogue Number71645030
Device Lot Number19GM20517
Initial Date Received by Manufacturer 01/29/2020
Supplement Date Received by Manufacturer01/29/2020
Initial Report FDA Received Date02/14/2020
Supplement Report FDA Received Date06/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-