EXACTECH, INC. OPTETRAK; CEMENTED FINNED TIB. TRA SZ 3F/4T Back to Search Results
Model Number 204-01-03
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Osteolysis (2377)
Date of Event 03/26/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This submission represents 1 of 163 events identified via active surveillance activities utilizing data form the (b)(6).Follow-ups for additional information cannot be performed for this complaint as the information was submitted as complete by the registry; there is no identification of facilities or surgeons.All available information about the event and/or patient has been provided and it is not possible to obtain any of the devices.Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported by (b)(6), this (b)(6) y/o, 170 cm, (b)(6) (bmi=28), male patient underwent primary total prosthetic replacement using cement of the right knee via medial parapatellar approach on (b)(6) 2008.The indication for the index procedure was osteoarthritis.On (b)(6) 2019, approximately 130 months post implantation, the patient underwent revision due to aseptic loosening of the tibia and osteolysis (lysistibia).Follow-ups for additional information cannot be performed for this complaint, information was submitted as complete by the (b)(6); there is no identification of facilities or surgeons.All available information about the event and/or patient has been provided.These devices are used for treatment not diagnosis.It is found in a review of the labeling and ifu 700-096-004, it is known that device specific risks consist of: fracture, migration, loosening, subluxation, or dislocation of the prothesis or any of its components, any of which may require a second surgical intervention or revision.Osteolysis is a known device specific risk, which may require a second surgical intervention or revision.As part of the preoperative assessment, the surgeon must ensure that no biological, biomechanical, or other factors exist that might adversely affect the surgery and/or postoperative period.Revisions or surgical interventions are a known complication found in joint replacements.All patients should be instructed on the limitations of the prosthesis and the possibility of subsequent surgery.The patient should be cautioned to monitor activities and protect the replaced joint from unreasonable stresses and follow the written instructions of the physician with respect to follow-up care and treatment.The surgeon must be fully knowledgeable about all aspects of the surgical technique to use the implants in accordance with the indications and contraindications and be trained according to the proper use of the system instrumentation and implants.It is a known complication that a patient¿s age, weight, or activity level would cause the surgeon to expect early failure of the system.It is known that bone mineral density at a total knee arthroplasty will decrease within several months post implant.Bone loss density can reach about 23% within one year of the post-op period.Values are dependent on the preoperative/post-surgical mechanical alignment, and the assessment method.Loss of the bone density post-op is found to be the result of the surgical procedure, peri-operative inflammation and bone remodeling associated with postoperative alterations to the mechanical load.[from: gallo, j.G.(2013, september).Osteolysis around total knee arthroplasty: a review of pathogenetic mechanisms.; retrieved from https://www.Ncbi.Nlm.Nih.Gov/pmc/articles/pmc4003873/].
 
Additional Manufacturer Narrative
Section h10: (h3) the evaluation noted that revision reported was likely the result of an insufficient bond between the implant and the bone, which led to aseptic (non-infected) loosening.However, this cannot be confirmed since the devices were not returned for evaluation.This submission represents 1 of 163 events identified via active surveillance activities utilizing data form the united kingdom national joint registry (uknjr).Follow-ups for additional information cannot be performed for this complaint as the information was submitted as complete by the registry; there is no identification of facilities or surgeons.All available information about the event and/or patient has been provided and it is not possible to obtain any of the devices.No information provided in the following section(s): a4, a5, and b6.The following section(s) have additional info: d2, d4 (catalog number, serial number, exp date, and udi number), g4, g5, g7, h1, h2, h3, h4, h6, and h7.Section h11: corrections made in the following section(s): (d2) common device name (d4) catalog number, serial number, exp date, and udi number (g5) 510k number (h4) manufacture date (h6) evaluation codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTETRAK
Common Device Name
CEMENTED FINNED TIB. TRA SZ 3F/4T
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
MDR Report Key9711088
Report Number1038671-2020-00159
Device Sequence Number19343009
Product Code JWH
UDI-Device Identifier10885862047588
UDI-Public10885862047588
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K932776
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/09/2012
Device Model Number204-01-03
Device Catalogue Number200-04-34
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/15/2020
Supplement Date Received by Manufacturer05/04/2020
Initial Report FDA Received Date02/14/2020
Supplement Report FDA Received Date05/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
Patient Weight80
-
-