| Model Number |
35615 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a tip fracture occurred.The target lesion was located in the liver.An accustick ii was selected for use in a percutaneous transhepatic biliary drainage (ptbd).During preparation, it was noted that the tip was broken.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that a tip fracture occurred.The target lesion was located in the liver.An accustick ii was selected for use in a percutaneous transhepatic biliary drainage (ptbd).During preparation, it was noted that the tip was broken.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device was returned for analysis.A visual inspection revealed that only accustick device was returned for analysis (cannula, dilator, sheath, introducer needle and stylet.).No visual issues were observed with the introducer needle and stylet.The devices did not have any broken section.Residues were observed over the dilator indicating use and handling of the device.The distal section of the cannula, dilator and sheath had a severe kink.The tip of the dilator was damaged.A functional inspection revealed that the components could not be taken apart due to severe kinking at distal section.
|
| |
|