| Model Number |
SSL-460 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/16/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported bilateral capsular contracture, baker grade iv.Device remains implanted.This is the right side record.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified; the weight the device within specification, deformation and creases fold.Cloudy in the gel observed after autoclave cycle.Based on the device analysis the final assessment is no issues found related with the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iv and "crease/folding of implant." device has been explanted.
|
| |
|
Event or Problem Description
|
|
Additionally, healthcare professional reported "crease/folding of implant." device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.
|
| |
|