Model Number 86-7428
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during a knee revision, the surgeon impacted the tibial stem/sleeve/tray construct.There was dense sclerotic bone in the canal.This caused the spline of the stem to bend and deform.He decided to bring the patient back and revised to a shorter cemented construct.Doi: (b)(6) 2020; dor: (b)(6) 2020; left knee.
 
Event or Problem Description
Update additional information received 21 feb 2020.It was reported that the reason for revision was due to the tibial tray that the spline on the component bent on.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned for evaluation.The investigation could not draw any conclusions about the reported event without the device to examine.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL STEM 115X14MM FLUTED
Common Device Name
UNIVERSAL KNEE STEMS AND SLEEVES : KNEE STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9711421
Report Number1818910-2020-04866
Device Sequence Number2513802
Product Code JWH
UDI-Device Identifier10603295217909
UDI-Public10603295217909
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K063633
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number86-7428
Device Catalogue Number867428
Device Lot NumberJ0177X
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/04/2020
Supplement Date Received by Manufacturer02/21/2020
07/20/2020
Initial Report FDA Received Date02/14/2020
Supplement Report FDA Received Date03/13/2020
07/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COMPETITOR CEMENT; MBT REVISION CEM TIB TRAY SZ 4; MBT TRAY SLEEVE POR M/L 45MM; TC3 RP TIBIAL INSERT S4,20.0
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-