• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AD-TECH MEDICAL INSTRUMENT CORP. CRANIAL DRILL BITS AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AD-TECH MEDICAL INSTRUMENT CORP. CRANIAL DRILL BITS AND ACCESSORIES Back to Search Results
Model Number DDK2-2.4-30X
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/05/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mdr 2183456-2020-00005 has been filed past the 30-day timeframe.Through the 2019 fda inspection, it was identified that while ad-tech was evaluating reportability for actual events in which death or serious injury had occurred, ad-tech had failed to submit events in which a malfunction was reported and could result in a serious injury if the event were to recur.As part of the investigation, a two (2) year retrospective review of complaints was performed to determine which complaints required submission of an mdr.Mdr-2183456-2020-00005 was submitted as a conservative measure due to the recalls tied to this issue.
 
Event or Problem Description
A user facility reported that the drill bit seized in the drill sleeve guide, possibly due to improper handling by the user facility.Back up tools were used and the case proceeded without patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRANIAL DRILL BITS AND ACCESSORIES
Common Device Name
DRILL BIT
Manufacturer (Section D)
AD-TECH MEDICAL INSTRUMENT CORP.
400 west oakview parkway
oak creek, wi
MDR Report Key9711519
Report Number2183456-2020-00005
Device Sequence Number2449545
Product Code HBE
UDI-Device Identifier00841823101167
UDI-Public(01)00841823101167(17)190401(10)0108085208140640
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K170442
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 02/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2019
Device Model NumberDDK2-2.4-30X
Device Catalogue NumberDDK2-2.4-30X
Device Lot Number208140640
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/05/2017
Initial Report FDA Received Date02/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2183456-08212019-R-004
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-