• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. OPTETRAK; PS TIBIAL INSERTS SZ 2, 9MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. OPTETRAK; PS TIBIAL INSERTS SZ 2, 9MM Back to Search Results
Model Number 204-01-02
Medical Device Problem Codes Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/08/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This submission represents 1 of 163 events identified via active surveillance activities utilizing data form the united kingdom national joint registry (uknjr).Follow-ups for additional information cannot be performed for this complaint as the information was submitted as complete by the registry; there is no identification of facilities or surgeons.All available information about the event and/or patient has been provided and it is not possible to obtain any of the devices.Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported by united kingdom national joint registry (uknjr), this (b)(6) y/o (bmi=25), male patient underwent primary total prosthetic replacement using cement of the left knee via medial parapatellar approach on (b)(6) 2009.The indication for the index procedure was osteoarthritis.The patient was implanted with a optetrak knee system - cemented posterior stabilised (ps) femoral component - size 2 (catalog 204-01-02, serial (b)(4)), optetrak knee system - cemented finned tibial tray - sizes 1f/3t 2f/3t (catalog 200-04-23, serial (b)(4)), optetrak knee system - posterior stabilised (ps) tibial insert - size 2 - 9mm (catalog 204-22-09, serial (b)(4)) and heraeus medical antibiotic loaded high viscosity bone cement (catalog 66017569, serial (b)(4)).On (b)(6) 2016, approximately 91 months post implantation, the patient underwent revision due to aseptic loosening femur; aseptic loosening tibia; wear of polyethylene component.The exactech were removed and replaced with nexgen rotating hinge femoral d left, rotating hinge knee precoat tibial plate size 4 and articular surface 17mm (fem size d).
 
Additional Manufacturer Narrative
The evaluation noted that revision reported was likely the result of a combination of prosthesis wear and aseptic loosening between the tibial and femoral components and the bone.The loosening may have led to an increase in cement and bone particles in the joint space and resulted in prosthesis wear or prosthesis wear may have contributed to particle-induced osteolysis.However, the contribution of loosening and prosthesis wear to the sequence of events cannot be determined as the devices were not available for evaluation.This submission represents 1 of 163 events identified via active surveillance activities utilizing data form the (b)(6).Follow-ups for additional information cannot be performed for this complaint as the information was submitted as complete by the registry; there is no identification of facilities or surgeons.All available information about the event and/or patient has been provided and it is not possible to obtain any of the devices.The following section(s) have additional info: d4 (catalog number, serial number, exp date, and udi number).Section h11: corrections made in the following section(s): (d2) common device name.(d4) catalog number, serial number, exp date, and udi number.(g5) 510k number.(h4) manufacture date.(h6) evaluation codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTETRAK
Common Device Name
PS TIBIAL INSERTS SZ 2, 9MM
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
MDR Report Key9711605
Report Number1038671-2020-00174
Device Sequence Number11272800
Product Code JWH
UDI-Device Identifier10885862047571
UDI-Public10885862047571
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/13/2011
Device Model Number204-01-02
Device Catalogue Number204-22-09
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/15/2020
Supplement Date Received by Manufacturer05/04/2020
Initial Report FDA Received Date02/14/2020
Supplement Report FDA Received Date05/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age47 YR
-
-