Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Information (3190)
Date of Event 12/06/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).It is unknown at this time if the device will be returned for evaluation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2020-00112.Medical products: tmj system right standard titanium mandibular component, part# 24-6545ti, lot# 703580a.Tmj system right fossa component, small, part# 24-6562, lot# 721330b.Unknown screws, part# unk, lot# unk.
 
Event or Problem Description
It was reported the patient underwent a revision of temporomandibular joint implants on the right side.The temporomandibular joint implants were removed and replaced with stock tmj prostheses.No additional patient consequences have been reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The complaint is confirmed as a revision surgery was reported.No product was returned as they were discarded; therefore, no functional tests or inspections could be performed.No x-rays, scans, pictures, or physician's reports were provided.The non-conformance database was reviewed for the mandible component; no non-conformances were found.There are no indications of manufacturing defects.This is the only complaint regarding heterotopic bone growth and dislocation for this part# 24-6545ti, lot# 703580a.For all non-custom tmj mandibular implants in the previous one year (from the notification date) regarding heterotopic bone growth, (b)(4).For all non-custom tmj mandibular implants in the previous one year (from the notification date) regarding dislocation, (b)(4).The most likely underlying cause of the complaint is a patient condition causing heterotopic bone growth.The cause of the dislocation could not be determined from the information provided, but it is possible that it is related to the bone growth.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
This follow-up report is being submitted to relay additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMJ SYSTEM RIGHT STANDARD TITANIUM MANDIBULAR COMPONENT
Common Device Name
JOINT, TEMPOROMANDIBULAR, IMPLANT
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key9711834
Report Number0001032347-2020-00111
Device Sequence Number3319723
Product Code LZD
UDI-Device Identifier00841036075118
UDI-Public00841036075118
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P020016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/23/2021
Device Model NumberN/A
Device Catalogue Number24-6545TI
Device Lot Number703580A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/16/2020
Supplement Date Received by Manufacturer07/27/2020
Initial Report FDA Received Date02/14/2020
Supplement Report FDA Received Date07/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age35 YR
-
-