LIVANOVA USA, INC. PROGRAMMING SOFTWARE Back to Search Results
Model Number 3000
Medical Device Problem Codes Computer Software Problem (1112); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/22/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a physician received a "current not being delivered" message that would go away after re-interrogating the patient multiple times.It was stated the generators were being interrogated after being at zero ma for an mri and is when the "current not delivered" pop-up message was being seen.After a system diagnostics was performed in both cases the message did not appear and the current was seen to be being delivered.Internal testing and data review identified that false low output current messages can occur when m3000 (b)(4) software programs a m103-106 generator after a specific programming sequence occurs.When m3000 (b)(4) tablets initiate a programming session with a m103-106 generator, where its output current is programmed off (0 ma), and then the output current is programmed back on to >0 ma, a low output current message would be seen when an in-session interrogation is performed.This would persist upon multiple in-session interrogations and if diagnostics weren't performed during the session, show on the session report.Running system diagnostics or ending and restarting the session will confirm functionality of the device, and resolve the ¿output current low¿ error message.The low output current messages in these cases are false and do not impact generator function.No further relevant information has been received to date.
 
Additional Manufacturer Narrative
Correction/removal reporting number: (b)(4).Voluntary medical device correction (remedial action) was initiated by the manufacturer on (b)(4) 2020, and the physician was provided a notification letter with recommended actions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Common Device Name
PROGRAMMING SOFTWARE
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
MDR Report Key9711839
Report Number1644487-2020-00257
Device Sequence Number2460678
Product Code LYJ
UDI-Device Identifier05425025750443
UDI-Public05425025750443
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 03/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3000
Was Device Available for Evaluation? No
Event Location Other
Initial Date Received by Manufacturer 01/22/2020
Supplement Date Received by Manufacturer02/21/2020
Initial Report FDA Received Date02/14/2020
Supplement Report FDA Received Date03/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age41 YR
-
-