This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on february 14, 2020.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 10, 11, 3331, 213, 22).Method code #1: 10 - testing of actual/suspected device.Method code #2: 11 - testing of device from same lot/batch retained by manufacturer.Method code #3: 3331 - analysis of production records.Results code: 213 - no device problem found.Conclusions code: 22 - known inherent risk of device.The returned sample was visually inspected in which no anomalies noted on the device.It was then set up on a drive motor and primed with a normal saline solution circuit.The rpm was set at 900 and ramped up by 300 rpm every ten minutes for one hour for a maximum of 2400 rpm.During each interval, the sample was observed for any running sound anomalies.No sound anomalies or abnormalities were noted during the test.A retention sample from the affected product code/lot number was obtained and tested the same way with no noted sound anomalies.The noise was not able to be replicated on the complaint sample; however, the issue experienced by the customer is likely due to an abnormal interaction between the seal rotor and stator surfaces.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|