| Model Number |
6207-31 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when opening the package and trying to take the device out during a laparoscopic colectomy, the needle was found to be disengaged.The product was not used to the patient and another device was used to complete case.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted one suture and one needle.The needle was returned disengaged from the suture.Upon microscopic inspection, instrument marks were noted near the swaged end of the needle.As no sealed samples were returned, a needle attachment test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.The needle should always be held in the middle where the diameter is greatest.Grasping the needle near the swaged end could cause bends or breaks, or damage the connection between the needle and suture.If information is provided in the future, a supplemental report will be issued.
|
| |
|