Model Number 10185P
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
This case was reviewed and investigated according to the manufacture¿s policy.It was reported during a diagnostic coronary procedure, inside the body, before normalization, the device was not recognized, no signal.Visual and microscopic inspection were performed on the returned device.The sensor was missing from the device.The probable cause for the separation could not be determined by the investigation.The manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.This product problem is being submitted because it could not be determined when the sensor separated from the manufacture¿s device.There is a potential for harm if the malfunction were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERRATA PLUS PRESSURE GUIDE WIRE, 185 CM STRAIGHT TIP
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
PHILIPS VOLCANO
2870 kilgore road
rancho cordova CA 95670
MDR Report Key9714641
Report Number2939520-2020-00005
Device Sequence Number408725
Product Code DQX
UDI-Device Identifier00845225002879
UDI-Public(01)00845225002879(17)220131(10)0301587667(90)989609001231
Combination Product (Y/N)Y
Initial Reporter CountryGM
PMA/510(K) Number
K161887
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/23/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model Number10185P
Device Catalogue Number400-0300.38
Device Lot Number0301587667
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2020
Initial Report FDA Received Date02/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-