From the device history record, the surgical towel pwtb04-stm/lot#20190802-23-sh was manufactured on 02nd aug 2019.No exception was recorded in the device history record that could lead to the reported incident.The average linting data is 0.207g / 10 pieces.No sample was available at the time of the investigation.According to supplier, the or towel is made of cotton, so cotton fiber is born and inevitable.Our supplier is continuously working with cardinal health to better control the linting and have implemented several measures to improve it: suctions machines have been installed in grey cloth rolling process, dyeing process and cutting process the suction process was added before product's final folding, and workers do it according to standard operation procedure requirement.Linting test method and acceptable criteria was stipulated to see the suction results.(=0.38g/10 pieces).In the folding process, supplier used one cloth pad under 100 pieces semi-finished products to avoid linting stuck onto the products during product's transfer.From the investigation, there is no abnormal situation that happened in production.Therefore, the root cause could not be determined.The complaint information was informed to the relevant sectors for their awareness.There is no action taken at this time, but we will continue to monitor the trend of this type of incident.
|
|
The customer reported excessive amount of lint coming off of the blue surgical towels pwtb04-stm from the neuro angiography pack sne40nacfa during a diagnostic angiogram.The towels were used to drape over the patient and layered to cover the drip lines.Several pauses were needed to clean the lint from gloves, catheters, instruments and saline bowl.There was no patient injury.No patient demographics were provided when inquired.Cardinal health is filing an mdr for potential risk to patient.
|