• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. OPTETRAK; CEMENTED FINNED TIB. TRA SZ 2F/2T

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. OPTETRAK; CEMENTED FINNED TIB. TRA SZ 2F/2T Back to Search Results
Model Number 204-01-02
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/04/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This submission represents 1 of 163 events identified via active surveillance activities utilizing data from (b)(6).Follow-ups for additional information cannot be performed for this complaint as the information was submitted as complete by the registry; there is no identification of facilities or surgeons.All available information about the event and/or patient has been provided and it is not possible to obtain any of the devices.Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported by (b)(6), this (b)(6), 162 cm, (b)(6) (bmi=30), female patient underwent primary total prosthetic replacement using cement of the right knee via medial parapatellar approach on (b)(6) 2011.The indication for the index procedure was osteoarthritis.The patient was implanted with a optetrak knee system - cemented posterior stabilised (ps) femoral component - size 2 (catalog 204-01-02, serial (b)(4)), optetrak knee system - cemented finned tibial tray - sizes 1f/2t 2f/2t (catalog 200-04-22, serial (b)(4)), optetrak knee system - posterior stabilised (ps) tibial insert - size 2 - 11mm (catalog 204-22-11, serial (b)(4)), optetrak knee system - three pegs patella - diameter 32mm (catalog 200-02-32, serial (b)(4)) and heraeus medical antibiotic loaded high viscosity bone cement (catalog 66017569, serial (b)(4)).On (b)(6) 2015, approximately 49 months post implantation, the patient underwent revision due to aseptic loosening tibia; instability.The exactech knee brand removed unavailable and replaced with acs« sc femoral component cemented size 3 right, acs« mb sc tibial component cemented size 3 and acs« mb sc pe- insert hyperflex size 3/10mm.
 
Additional Manufacturer Narrative
Section h10: (h3) the evaluation noted that revision reported was likely the result of an insufficient bond between the implant and the bone, which led to aseptic (non-infected) loosening, possibly secondary to increased biomechanical stress and instability.However, the contributions of each factor to the sequence of events cannot be confirmed since the devices were not returned for evaluation and no additional first hand details were provided.This submission represents 1 of 163 events identified via active surveillance activities utilizing data form the united kingdom national joint registry (uknjr).Follow-ups for additional information cannot be performed for this complaint as the information was submitted as complete by the registry; there is no identification of facilities or surgeons.All available information about the event and/or patient has been provided and it is not possible to obtain any of the devices no information provided in the following section(s): a4, a5, and b6.The following section(s) have additional info: d2, d4 (catalog number, serial number, exp date, and udi number), g4, g5, g7, h1, h2, h3, h4, h6, and h7.Section h11: corrections made in the following section(s): (d2) common device name.(d4) catalog number, serial number, exp date, and udi number.(g5) 510k number.(h4) manufacture date.(h6) evaluation codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTETRAK
Common Device Name
CEMENTED FINNED TIB. TRA SZ 2F/2T
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
MDR Report Key9714820
Report Number1038671-2020-00197
Device Sequence Number11336655
Product Code JWH
UDI-Device Identifier10885862047571
UDI-Public10885862047571
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K932776
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2015
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/28/2015
Device Model Number204-01-02
Device Catalogue Number200-04-22
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/15/2020
Supplement Date Received by Manufacturer05/04/2020
Initial Report FDA Received Date02/14/2020
Supplement Report FDA Received Date05/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age57 YR
Patient Weight79
-
-