|
Model Number 2426-0500 |
Device Problems
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
Patient Problems
Low Blood Pressure/ Hypotension (1914); Low Oxygen Saturation (2477)
|
Event Date 02/02/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Although requested, device has not been received.A follow up report will be submitted with failure investigation results should the device be received for evaluation.
|
|
Event Description
|
It was reported that the infusion pumps were transferred to iv poles and patient was transported computed tomography (ct) scan department.While patient was in ct scan, the blood pressure significantly dropped to 68 mmhg despite titrating the levophed infusion from 0.2mcg/kg/min to 0.95mcg/kg/min.Patient's blood pressure was stable all day prior to the event.Upon inspection, a small hole "pinhole" was found in the tubing just below the pump segment and the levophed solution was leaking onto the floor.The clinicians further noted that the drop in blood pressure was also attributed to derecruitment of patient's lung regions (i.E., airway and alveolar closure) when the patient was taken off the ambu bag (bag valve mask device) and switched to mechanical ventilator, because the patient also had a drop in oxygen saturation (spo2) that required interventions.The rn called the physician over the phone and immediately provided verbal order for epinephrine iv, which was given to the patient after being obtained from the code cart.The patient's blood pressure recovered, and the physician immediately came down to the ct scan room to intervene as well.The patient was brought back to trauma and neurosurgery intensive care unit (tnicu) for further management.Patient's systolic blood pressure and oxygen saturation recovered.The charge rn, resource nurse and icu fellow were informed.
|
|
Manufacturer Narrative
|
The customer¿s report that the hole ¿cut/slice in tubing and leaked" was confirmed.Visual inspection observed a cut/slice on the tubing just below the second smartsite outlet port below the lower fitment.During functional testing the set leaked from the tubing cut that was observed during visual inspection.The root cause of the hole ¿cut¿ was not identified.
|
|
Event Description
|
It was reported that the infusion pumps were transferred to iv poles and patient was transported computed tomography (ct) scan department.While patient was in ct scan, the blood pressure significantly dropped to 68 mmhg despite titrating the levophed in 0.9% normal saline 500ml infusion from 0.2mcg/kg/min to 0.95mcg/kg/min.Patient's blood pressure was stable all day prior to the event.Upon inspection, a small hole "pinhole" was found in the tubing just below the pump segment and the levophed solution was leaking onto the floor.The leak was not noticed until the nurse investigated why the patient was not responding to the infusion.The clinicians further noted that the drop in blood pressure was also attributed to derecruitment of patient's lung regions (i.E., airway and alveolar closure) when the patient was taken off the ambu bag (bag valve mask device) and switched to mechanical ventilator, because the patient also had a drop in oxygen saturation (spo2) that required interventions.The rn called the physician over the phone and immediately provided verbal order for epinephrine iv, which was given to the patient after being obtained from the code cart.The patient's blood pressure recovered, and the physician immediately came down to the ct scan room to intervene as well.The patient was brought back to trauma and neurosurgery intensive care unit (tnicu) for further management.Patient's systolic blood pressure and oxygen saturation recovered.The charge rn, resource nurse and icu fellow were informed.
|
|
Search Alerts/Recalls
|
|
|