• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL HEAL COLLAR 3.5X4.5, 5MM; HEALING ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL HEAL COLLAR 3.5X4.5, 5MM; HEALING ABUTMENT Back to Search Results
Model Number HC345
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Weight unknown / not provided.
 
Event or Problem Description
It was indicated that the healing collar failed to be placed.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).One healing collar (hc345) was returned for investigation.Visual evaluation of the as returned product identified signs of wear due to usage around the threads.Functional testing was performed using an in-house implant.The device could not seat as normal on the implant.Pre-existing condition noted on the per was high bone density ¿ type i.Doctor was attempting to place the device on an unknown tooth location when the incident occurred.X-ray or picture images were not provided and no other information was provided.Dhr review for the lot (2019011477) had revealed no deviations nor non-conformances which could have caused or contributed to the reported event.All products were conforming at the time they left zimmer biomet.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the lot (2019011477) for similar event and no other complaint was identified.Therefore, based on the available information, device malfunction has occurred and the reported event was confirmed.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEAL COLLAR 3.5X4.5, 5MM
Common Device Name
HEALING ABUTMENT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key9715942
Report Number0002023141-2020-00315
Device Sequence Number16231477
Product Code NHA
UDI-Device Identifier00889024015005
UDI-Public(01)00889024015005(17)240128(10)2019011477(241)HC345
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K111852
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2024
Device Model NumberHC345
Device Catalogue NumberHC345
Device Lot Number2019011477
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/23/2020
Supplement Date Received by Manufacturer04/09/2020
Initial Report FDA Received Date02/17/2020
Supplement Report FDA Received Date04/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age25 YR
-
-