| Model Number |
9733068 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was unavailable from the site.The biopsy needle kit was returned to the manufacturer for evaluation.Testing found that the depth stop on the needle was not drilled through the set screw.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used for a cranial biopsy procedure.It was reported that the screw on the adjustable depth collar would not tighten.The biopsy needle was replaced and the issue was resolved.There was no reported delay and no reported impact to patient outcome.
|
| |
|