Model Number N/A
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/11/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.(b)(6).Pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported an ultrathane mac-loc locking loop multipurpose drainage catheter was inspected prior to use.A hair was noticed in the primary packaging of the product.No other adverse effects were reported for this incident.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation.Brigham & women's hospital (united states) informed cook that they noticed a hair inside the packaging of an ultrathane mac-loc locking loop multipurpose drainage catheter.The device was not used.A review of the complaint history, device history record, instructions for use (ifu), quality control, were conducted during the investigation.The customer did not return the device and did not provide photos of the failure.Therefore, no visual inspection was performed.Additionally, a document-based investigation evaluation was performed.There are adequate inspection activities in place to detect foreign matter prior to distribution.The device history record (dhr) was reviewed.The lot revealed no nonconformances.A complaint database search was completed.No additional complaints were found for the lot.Based on this information, there is no evidence of nonconforming material in house or in the field.The instructions for use provided with this lot [t_multi_rev5] state: "sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile." because the device was not returned and no photos were provided, cook cannot confirm the device was nonconforming.Based on the information provided, no examination of returned product, and the results of the investigation, it is possible that a manufacturing or quality control deficiency caused the failure.The appropriate personnel have been notified.Per the quality engineering risk assessment, no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER
Common Device Name
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9716992
Report Number1820334-2020-00368
Device Sequence Number16234765
Product Code GBO
UDI-Device Identifier00827002097667
UDI-Public(01)00827002097667(17)221106(10)10134123
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date11/06/2022
Device Model NumberN/A
Device Catalogue NumberULT10.2-38-45-P-6S-CLM-RH
Device Lot Number10134123
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/11/2020
Supplement Date Received by Manufacturer04/27/2020
Initial Report FDA Received Date02/17/2020
Supplement Report FDA Received Date05/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-