| Model Number |
DLXPLE38 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Pain (1994); Loss of Range of Motion (2032)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This event will be evaluated once the investigation is complete.Trends will be evaluated.
|
| |
|
Event or Problem Description
|
|
Allegedly, patient was revised due to elevated cobalt ion levels, pain and stiffens of the hip.It was discovered that the device failed due to corrosion of the oblong taper of modular neck where it seated in the pocked of the titanium stem.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated lot information.
|
| |
|
Search Alerts/Recalls
|