| Model Number |
00-1050-6004 |
| Medical Device Problem Codes |
Fracture (1260); Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reference: (b)(4).Concomitant medical products: unknown locking screw: item number; unknown, lot number; unknown qty of four (4).Customer has indicated that the product will be returned to orthopediatrics for investigation.Multiple mdr reports were filed for this event, please see associated reports: 3006460162-2020-00032, 3006460162-2020-00033, 3006460162-2020-00034 and 3006460162-2020-00035.
|
| |
|
Event or Problem Description
|
|
It was reported that the dfos plate failed and the screws broke.The patient was revised.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: h6 patient code: 1924, h6 device code: 1260, h6 method code: 3331 - 4109, h6 result code: 3252, h6 conclusion code: 4315, d10 device available for evaluation: yes.Complaint sample was evaluated and the reported event was confirmed.Product evaluation confirmed broken screws.The plate was found to be undamaged.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no were trends identified.Root cause was unable to be determined.
|
| |
|