A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to (b)(4) has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes. device evaluation: visual analysis of the returned device identified: wear abrasion, opening on anterior, crease flat.Leak test and microscopic analysis was performed which identified: striated opening, sharp opening and observed void >1 mm in non-textured, valve-to shell-bond area.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a striated opening assessed as a surgical damage consistent in the use of some surgical tool.A sharp opening on valve bond edge assessed as an unidentified (tear) opening.Additional, changed, and/or corrected data: b.5, d.7., d.10., h.3, h.6.
|