| Model Number |
PCB00 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.If explanted, give date: not applicable, as the lens remains implanted.(b)(6).Device evaluation: product testing could not be performed as the product was not returned for evaluation.Based on the information provided the lens remains implanted.Therefore, the reported issue could not be verified, and product quality deficiency could not be verified.Video is available and was evaluated.Visible rings on the lens is possible.As part of the manufacturing process the lenses are inspected and discarded if the iol exceeds the established acceptance criteria defined in the manufacturing process.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaint was received from this production order.There was no discrepancy and/or deviation found during the mrr.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that clearly 1 visible groove on the periphery on the optic of the lens could be seen during implantation.No explant is planned and no impact on patient vision has been reported.Additional information received states no interventions are planned, outcome is good and no reports of halos or other visual issues.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|