• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH WORKING INSERT, ALBARRAN, TWO WAY; WORKING INSERTS WITH ALBARAN LEVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH WORKING INSERT, ALBARRAN, TWO WAY; WORKING INSERTS WITH ALBARAN LEVER Back to Search Results
Model Number A20972A
Medical Device Problem Codes Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that it was noticed that the albarran lever at the distal end of the working insert could not be raised anymore and that its fixing pin was missing.It is unknown when the fixing pin fell off and whether the fixing pin possibly fell into the patient¿s body cavity.However, it was suspected that it would have been flushed out from the patient with irrigation fluid if it had detached during a procedure.No further information was provided and there was no report about an adverse event or patient injury.
 
Additional Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation but to olympus medical systems corporation (omsc), japan (returned to omsc on (b)(6) 2020).The evaluation/investigation confirmed that the fixing pin of the albarran lever is missing.Furthermore, the distal end of the working element appears discolored due to corrosion.The cause of this damage is most likely wear and tear.Therefore, this event/incident was attributed to use error.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the working insert without showing any abnormalities.The case will be closed from olympus side with no further actions but the reported event/incident will be recorded for trending and surveillance purposes.Furthermore, the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WORKING INSERT, ALBARRAN, TWO WAY
Common Device Name
WORKING INSERTS WITH ALBARAN LEVER
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key9719712
Report Number9610773-2020-00070
Device Sequence Number11675119
Product Code GCP
UDI-Device Identifier04042761020374
UDI-Public04042761020374
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA20972A
Device Catalogue NumberA20972A
Device Lot Number125W
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer02/21/2020
Initial Report FDA Received Date02/18/2020
Supplement Report FDA Received Date03/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-