• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. X-ALT HIP; LINER/NON-HOODED-NEU, MP5, HXE-PLUS, 28MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. X-ALT HIP; LINER/NON-HOODED-NEU, MP5, HXE-PLUS, 28MM Back to Search Results
Model Number 931-28-748
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/20/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 3.8 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery: patient had an infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-ALT HIP
Common Device Name
LINER/NON-HOODED-NEU, MP5, HXE-PLUS, 28MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
MDR Report Key9720007
Report Number1644408-2020-00091
Device Sequence Number19433324
Product Code OQG
UDI-Device Identifier00888912099431
UDI-Public(01)00888912099431
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130365
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/28/2019
Device Model Number931-28-748
Device Catalogue Number931-28-748
Device Lot Number758N1901
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2020
Initial Report FDA Received Date02/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-281 LOT 857B1027
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
-
-