It was reported that a procedure was performed to treat a lesion in the left superficial femoral artery.
A command 18 guide wire was advanced with a 5x200 armada 18 percutaneous transluminal angioplasty (pta) catheter without resistance.
The armada was inflated, and the balloon folds were bunched after deflation.
The guide wire and pta faced resistance with each other during removal, so the devices were removed together.
The polymer of the command guide wire was peeled.
It was confirmed that no portion of the devices remain in the anatomy.
An unspecified command guide wire and unspecified armada pta were used to successfully complete the procedure.
There were no adverse patient effects and no clinically significant delay in the procedure.
No additional information was provided.
|