| Model Number |
331641 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: description of issue: on (b)(6) 2020, patient received a femoral nerve block.On (b)(6) 2020, the catheter was removed and 1 cm of the blue tip was noted to be missing.It appeared that the tip was sheared off.An x-ray was done and no catheter was seen.No redness, tenderness, swelling was noted at the site.Surgeon was consulted and it was recommended to monitor the site.No additional treatments were performed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).One (1) used catheter without packaging was returned for evaluation.Visual evaluation of the sample showed the catheter to be torn/sheared near the tip.Based on the evaluation results, this type of defect could have happen during application.User should refer to ifu which states, "never pull the catheter back through the epidural needle due to the possibility of shearing or kinking the catheter.When removing the catheter, excessive force should never be applied.If the catheter becomes difficult to withdraw, consult standard textbooks for specific techniques." we will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|