| Model Number |
4P54-25 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a patient information: no specific patient information was provided.
|
| |
|
Event or Problem Description
|
|
The customer obtained a false positive architect hbsag qualitative confirmatory result.Sample id (b)(6) generated repeat reactive hbsag results of 16.82 s/co with a similar retest and confirmed reactive with the hbsag confirmatory assay.A new specimen (s2676581) was collected two weeks later and generated repeat reactive hbsag results 16.27 s/co with a similar retest result and not confirmed when tested with the hbsag confirmatory assay.The sample was also sent to an alternate labe and generated a non-reactive result.A third sample (t4178124) generated non-reactive results on confirmation.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of trending reports did not identify any related trends for the architect hbsag qualitative confirmatory assay.Normal complaint activity was identified for reagent lot 08014fn00.Testing of panels which mimic patient samples was performed using a retained kit of lot 08014fn00.All specifications were met indicating that the lot is performing acceptably.Manufacturing documentation for the reagent lot was reviewed and did not identify any issues associated with the issue.Labeling was reviewed and sufficiently addresses the customer's issue.No systemic issue or deficiency of the architect hbsag qualitative confirmatory assay was identified.
|
| |
|
Search Alerts/Recalls
|